Changes in the biomechanical properties of articular cartilage are one of the first signs of the tissue degeneration. We have developed a small size indentation instrument for the quantification of cartilage stiffness under arthroscopic control. During measurement, the indenter imposes a constant deformation on the cartilage and the maximal indenter force, by which the cartilage resists the deformation, is used as a measure for cartilage stiffness. The instrument has been tested in laboratory conditions with elastomer and cadaver knee joint cartilage samples. A linear relationship was found between indenter force and elastomer stiffness (r = 0.990, n = 14) as well as between indenter force and cartilage shear modulus obtained by a reference device (r = 0.879, n = 22). Also, the correlation between two repeated measurements at the measurement sites, used to evaluate the reproducibility, was linear (r = 0.953, n = 16). Quantitative detection of cartilage stiffness is possible with the instrument.
No takes yet. Share an insight, caveat, or question.
Lyyra et al. (1995) studied this question.