Why the study?
Is the combined use of a transvenous pacemaker and an extravascular implantable cardioverter-defibrillator safe and feasible without oversensing?
Is the combined use of a transvenous pacemaker and an extravascular implantable cardioverter-defibrillator safe and feasible without oversensing?
Answers are grounded in this paper and cite their sources.
Simultaneous use of an extravascular ICD and a transvenous pacemaker is safe and feasible without oversensing issues, provided devices are appropriately programmed.
1)When combining two functional device in one patient always check for crosstalk in normal setting but also in the worsted case scenario. This to avoid potential problems that might occur during life time of the devices or leads. You need to be prepared for those possible issues2)It is always possible that after a implant of a new type of device a new indication occurs for that patient. Just explanting and implanting a new device that fits both indication is not always necessary. Be creative in a safe way and see what is already done in the literature for possible options or consult expert centers for advice.3)Consult expert centers or company staff for advice. They might have already encountered such a problem and may be able to assist with trouble shooting
No takes yet. Share an insight, caveat, or question.
Römers et al. (2023) studied this question.