Why the study?
Does a steroid eluting epicardial plaque electrode improve pacing thresholds and sensing compared to steroid-free and standard leads in a canine model?
Does a steroid eluting epicardial plaque electrode improve pacing thresholds and sensing compared to steroid-free and standard leads in a canine model?
A novel steroid-eluting epicardial electrode provides superior pacing thresholds and sensing characteristics in a canine model, supporting further development.
Answers are grounded in this paper and cite their sources.
Supports preclinical development of steroid-eluting epicardial electrodes; leaves open translation to clinical pacing practice.
Acute and chronic canine atrial and ventricular thresholds and sensing were compared for a steroid eluting epicardial plaque electrode, a similar steroid free plaque electrode and two standard leads. The steroid eluting electrode had low, stable thresbolds in atrium and ventricle compared to the steroid free plaque and standard leads. This new electrode also provided improved atrial sensing. Thus, the concept of a steroid eluting epicardial electrode shows promise for further evaluation.
No takes yet. Share an insight, caveat, or question.
KENNETH B. STOKES (1988) studied this question.