Why the study?
Do steroid-eluting unipolar electrodes maintain low pacing thresholds in patients requiring ventricular pacing?
Do steroid-eluting unipolar electrodes maintain low pacing thresholds in patients requiring ventricular pacing?
Steroid-eluting unipolar electrodes maintain very low pacing thresholds in both the early and chronic follow-up phases without loss of sensing.
Answers are grounded in this paper and cite their sources.
May support steroid-eluting electrodes for low thresholds in ventricular pacing; hypothesis-generating and should not change practice.
In continuing search of low chronic threshold leads, a new concept of electrode design which is capable of delivering corticosteroids at the myocardial tissue interface has been made available by Medtronic. Twenty-three patients, 17 females and 6 males, were either implanted with 4003 (n = 21) or 5023 (n = 2) steroid-eluting electrodes in the ventricular chamber. Pacing modes utilized were VVIM (n = 13) or DDD (n = 10). Pulse generators used were Medtronic (7005, 8317, 8329) Pacesetter (285) and Intermedics (283). Thresholds at the time of implantation at 0.50 msec pulse width were 0.40 +/- 0.02 volts at 0.66 +/- 0.05 milliamps. Resistance and R wave measured were 565.43 +/- 22.07 ohms and 9.24 +/- 1.06 mv, respectively. Chronic thresholds were checked on routine follow-up visits by either decreasing pulse width and/or pulse amplitude. Data is being reported between 1 and 88 (23.22 +/- 4.35) weeks. Pulse width threshold at 2.5 volts were 0.10 msec (n = 11) and 0.05 msec or lower (n = 12). At 5.0 volts no loss of capture was seen at 0.05 msec (n = 22) except in one patient at 0.10 msec. Pulse width thresholds in the first 24 weeks were lower than 0.20 msec at 2.5 volts (n = 15) and less than 0.70 msec. at 0.8 volts (n = 6). No loss of sensing was seen by electrocardiographic analysis at the time of threshold checks with the pulse generator at standard setting of the R wave. Thus, in this initial report, the steroid-eluting electrodes have demonstrated very low thresholds both in the early and chronic follow-up phase.(ABSTRACT TRUNCATED AT 250 WORDS)
No takes yet. Share an insight, caveat, or question.
Pirzada et al. (1988) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: