Steroid-eluting ventricular pacing electrodes demonstrate a significant increase in pacing threshold between 18 and 36 months post-implantation, but remain stable up to 5 years.
Answers are grounded in this paper and cite their sources.
Threshold rise at 18-36 months may warrant closer surveillance; leaves open long-term durability of steroid-eluting leads in larger cohorts.
Sixteen patients with Medtronic 4003 steroid-eluting electrodes implanted in the ventricular position were followed over 5 years. In each patient a special type of Medtronic 2443 pacemaker was implanted to allow programming of output at 1.35 V. Chronic threshold values in these patients measured at an output of 1.35 V were stable over the first 18 months of follow-up. Mean values were: 0.06 +/- 0.03 msec at 6 months and 0.08 +/- 0.02 msec at 18 months; these did not differ from each other significantly. However, during the period from 18 to 36 months postimplantation, a significant increase in mean pacing threshold was observed: 0.08 +/- 0.02 msec at 18 months postimplantation versus 0.14 +/- 0.05 msec at 36 months (P < 0.01). After 36 months, the chronic pacing threshold remained stable until the end of the 5-year follow-up period. Further long-term study of chronic threshold behavior of steroid-eluting electrodes measured at low amplitudes is warranted.