Why the study?
Does the ultrasonic cardiac output monitor (USCOM) accurately measure stroke volume changes compared to the oesophageal Doppler monitor (ODM) in patients undergoing major abdominal surgery?
Does the ultrasonic cardiac output monitor (USCOM) accurately measure stroke volume changes compared to the oesophageal Doppler monitor (ODM) in patients undergoing major abdominal surgery?
USCOM shows good concordance and accurately discriminates clinically significant changes in stroke volume compared to ODM during major abdominal surgery.
Answers are grounded in this paper and cite their sources.
May not yet replace oesophageal Doppler for perioperative monitoring; leaves open USCOM utility in larger validation studies.
Background Perioperative interventions, targeted to increase global blood flow defined by explicit measured goals, reduce postoperative complications. Consequently, reliable non-invasive estimation of the cardiac output could have far-reaching benefit. Methods This study compared a non-invasive Doppler device – the ultrasonic cardiac output monitor (USCOM) – with the oesophageal Doppler monitor (ODM), on 25 patients during major abdominal surgery. Stroke volume was determined by USCOM (SV USCOM ) and ODM (SV ODM ) pre and post fluid challenges. Results A ≥ 10% change (Δ) SV USCOM had a sensitivity of 94% and specificity of 88% to detect a ≥ 10% Δ SV ODM ; the area under the receiver operating curve was 0.94 (95% CI 0.90–0.99). Concordance was 98%, using an exclusion zone of <10% Δ SV ODM . 135 measurements gave median SV USCOM 80 ml (interquartile range 65–93 ml) and SV ODM 86 ml (69–100 ml); mean bias was 5.9 ml (limits of agreement −20 to +30 ml) and percentage error 30%. Conclusions Following fluid challenges SV USCOM showed good concordance and accurately discriminated a change ≥10% in SV ODM .
No takes yet. Share an insight, caveat, or question.
Hodgson et al. (2015) studied this question.