Key Points
- This study aims to assess the accuracy of measuring intraventricular volume using conductance catheter technique in isolated canine hearts.
- Left ventricular volumes were measured using a validated servosystem with an electroconductive balloon.
- Hearts were simulated to eject and fill physiologically with a computer-simulated arterial loading system.
- The relationship between balloon volume (Vbal) and conductance volume (Vcath) was analyzed, considering influenced right ventricular volumes.
- There was a strong linear correlation between Vbal and Vcath with r² = 0.982 ± 0.014.
- The equation was established: Vcath = 0.82 (± 0.05) Vbal + 26.7 (± 11.8) ml.
- Right ventricular volume had minimal effect, shifting the relationship by less than 3 ml.
Structured PICO
PPopulationSix isolated, ejecting, canine left ventricles
IInterventionIntraventricular volume measurement by intracavitary electrical conductance (conductance catheter method)
CComparatorDirect volume measurement by a validated servosystem using an electroconductive balloon (Vbal)
OOutcomeAccuracy and linearity of volume measurement (relationship between Vbal and Vcath)surrogate
The conductance catheter method accurately estimates left ventricular volume changes in isolated canine hearts, providing a signal proportional to actual volume.