Conductance catheter tracks right ventricular stroke volume closely in preclinical testing.
Why the study?
The geometry of the right ventricle makes measurement of its volume by intraventricular fluid conductance more problematical than for the left ventricle.
Does an eight-electrode conductance catheter accurately measure right ventricular volume changes compared to an electromagnetic blood flowmeter in anesthetized greyhounds?
Does an eight-electrode conductance catheter accurately measure right ventricular volume changes compared to an electromagnetic blood flowmeter in anesthetized greyhounds?
Catheter measurement of conductance provides a highly correlated estimation of right ventricular stroke volume compared to electromagnetic flowmetry in an animal model.
Answers are grounded in this paper and cite their sources.
Supports conductance catheter for RV volumes in dogs; leaves open human validation before clinical use.
The electrical conductivity of blood is sufficiently higher than that of myocardium to make feasible the detection of cardiac volume changes by measurement of intraventricular fluid conductance. An eight-electrode catheter was used to inject an alternating current (100 microA or less, at 1500 Hz) via the two electrodes nearest the ventricular base and apex, then the resulting five voltage differences between adjacent pairs of the six intervening electrodes were measured. When current amplitude was held constant, the cross-sectional area of the ventricular cavity slice defined by planes perpendicular to the catheter through the relevant pair of electrodes was inversely proportional (to the first order) to the voltage difference. Measurement of multiple segments compensated for isovolumic cavity shape changes. The technique had previously been shown to measure left ventricular volume successfully, but the geometry of the right ventricle made this measurement more problematical. Using open-chested, anesthetized greyhounds, we compared the catheter-measured right ventricular volume change with stroke volume as measured by a pulmonary arterial electromagnetic blood flowmeter. With optimal catheter placement, good correlation between stroke volume and catheter-measured volume changes was achieved when stroke volume was perturbed on a beat-to-beat basis. In six data records from three dogs, involving two different means of varying stroke volume (rapid injection of blood and sinus node irritation), the correlations yielded r2 values between 0.82 and 0.98.(ABSTRACT TRUNCATED AT 250 WORDS)
No takes yet. Share an insight, caveat, or question.
Woodard et al. (1987) studied Right ventricular volume measurement (n=3). Catheter measurement of intraventricular fluid conductance vs. Pulmonary arterial electromagnetic blood flowmeter was evaluated on Correlation between catheter-measured right ventricular volume change and stroke volume (r2 values between 0.82 and 0.98). Catheter measurement of intraventricular fluid conductance correlated well with stroke volume measured by an electromagnetic flowmeter in greyhounds.