The Effect of Radiofrequency Catheter Ablation on Permanent Pacemakers: An Experimental Study
Why the study?
Does transcatheter radiofrequency ablation cause electromagnetic interference and malfunction in permanent pacemakers?
Population
19 pulse generators (16 models from 7 manufacturers) acutely implanted in 12 dogs
Design
Preclinical
Follow-up
Acute (during and after ablation)
Authors
Loading...
Radiofrequency ablation performed within 1 cm of a permanent pacing lead can induce severe pacemaker malfunctions, including runaway and false inhibition, highlighting the need for backup pacing and careful monitoring.
Does transcatheter radiofrequency ablation cause electromagnetic interference and malfunction in permanent pacemakers?
Radiofrequency ablation performed within 1 cm of a permanent pacing lead can induce severe pacemaker malfunctions, including runaway and false inhibition, highlighting the need for backup pacing and careful monitoring.
Chin et al. (1990) studied this question.