May 1, 1995Pacing and Clinical Electrophysiology

Pacemaker Function During Radiofrequency Ablation

View Full Paper
Ask AI
Bookmark
Share

Why the study?

How do pacemakers behave during radiofrequency ablation procedures?

Population

25 patients with antitachycardia or antibradycardia pacemakers undergoing radiofrequency ablation

Design

Case_series

Key result

Radiofrequency ablation in patients with pacemakers resulted in sensing failures in 32.0% and pacing failures in 16.0%, with no pacemaker damage observed.

Authors

DPD. PfeifferJTJürgen TebbenjohannsBSBurghard Schumacher

Discussion

Loading...

Member takes

Overview

Transient failures warrant periprocedural vigilance in paced patients; case reports leave open optimal device protocols.

Study Design

Type

Observational (n=25)

Structured PICO

How do pacemakers behave during radiofrequency ablation procedures?

P
Population
25 patients with 13 different pacemaker devices undergoing radiofrequency ablation.
E
Exposure
Radiofrequency ablation
O
Outcome
Pacemaker reaction/behavior during and after ablation proceduressafety

Radiofrequency ablation can cause transient sensing and pacing failures in patients with pacemakers, highlighting the need for generator deactivation and external pacing during the procedure.

Cite This Study

Pfeiffer et al. (1995) conducted an observational in Patients with pacemakers undergoing radiofrequency ablation (n=25). Radiofrequency ablation was evaluated on Pacemaker reaction (sensing and pacing failures). Radiofrequency ablation in patients with pacemakers resulted in sensing failures in 32.0% and pacing failures in 16.0%, with no pacemaker damage observed.