March 23, 2004AnesthesiologyOpen Access

Extracardiac Radiofrequency Ablation Interferes with Pacemaker Function but Does Not Damage the Device

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Does extracardiac radiofrequency ablation cause electromagnetic interference or damage in patients with a permanent pacemaker?

Population

1 patient with a right adrenal metastasis and high-grade suprahisian block with an implanted permanent…

Design

Case_report

Follow-up

Immediate postoperative period

Key result

Extracardiac radiofrequency ablation induced electromagnetic interference that temporarily inhibited the implanted pulse generator and caused irregular pacing, but did not permanently damage it.

Authors

NTNg Yuet TongHRHsieh Jing RuHLHui Ling

Discussion

Loading...

Member takes

Overview

Transient EMI risk warrants vigilance near pacemakers; case report leaves open standardized perioperative protocols.

Study Design

Type

Case Report (n=1)

Structured PICO

Does extracardiac radiofrequency ablation cause electromagnetic interference or damage in patients with a permanent pacemaker?

P
Population
One male patient with a permanent dual-chamber pacemaker undergoing extracardiac radiofrequency ablation for a right adrenal metastasis.
E
Exposure
Extracardiac radiofrequency ablation (four 12-min ablations, max 200 W, 480 kHz) of a right adrenal metastasis.
O
Outcome
Pacemaker function and electromagnetic interference during and after radiofrequency ablation.safety

Extracardiac radiofrequency ablation in close proximity to a permanent pacemaker can cause significant but transient electromagnetic interference without permanently damaging the device.

Limitations

  • Limited information obtained from reported cases
  • Limited information from reported cases

Cite This Study

Tong et al. (2004) conducted a case report in Right adrenal metastasis with permanent pacemaker (n=1). Extracardiac radiofrequency ablation was evaluated on Pacemaker function and electromagnetic interference. Extracardiac radiofrequency ablation induced electromagnetic interference that temporarily inhibited the implanted pulse generator and caused irregular pacing, but did not permanently damage it.

synapsesocial.com/papers/6a847a49c0e58a00207bfd30https://doi.org/10.1097/00000542-200404000-00053
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1The Effect of Radiofrequency Catheter Ablation on Permanent Pacemakers: An Experimental Study1990 · 63 citations
  2. 2Acute Effects of Radiofrequency Ablation of Atrial Arrhythmias on Implanted Permanent Pacing Systems1996 · 51 citations
  3. 3Pacemaker Function During Radiofrequency Ablation1995 · 52 citations
  4. 4Radiofrequency Catheter Atrioventricular Node Ablation in Patients with Permanent Cardiac Pacing Systems1994 · 40 citations
  5. 5Safety of Pacemaker Implantation Prior to Radiofrequency Ablation of Atrioventricular Junction in a Single Session Procedure2000 · 25 citations