Why the study?
Modern permanent pacemakers have unique features that can cause inappropriate pacing inhibition leading to symptoms, which can be difficult to identify when intermittent and hard to reproduce.
This case highlights a rare cause of inappropriate ventricular pacing inhibition due to a specific pacemaker electrogram storage feature, which can be resolved by adjusting the ventricular lead sensitivity.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to interrogate for intermittent inhibition in unexplained PPM symptoms; leaves open systematic evaluation of algorithm safety.
BACKGROUND: Modern permanent pacemakers (PPMs) have individual features designed to identify cardiac rhythm abnormalities and improve their performance. Inappropriate pacing inhibition may be an undesired outcome from these features and cause symptoms in patients who require frequent pacing, leading to dizziness, and syncope. Inappropriate inhibition can be difficult to identify in circumstances that are intermittent and difficult to reproduce. CASE SUMMARY: A 57-year-old female underwent a mitral valve replacement (MVR) for severe mitral stenosis. One month following MVR, she presented with symptomatic third-degree atrioventricular block, and a dual-chamber PPM (Advisa™, Medtronic, Minneapolis, USA) was implanted and programmed DDD 50-130 b.p.m. At the 3-month follow-up, she reported frequent episodes of lightheadedness. She was found to have intermittent ventricular pacing inhibition on a 48-h Holter monitor due to an internal function of the Advisa™ series of PPMs that attempts to store an electrogram (EGM) every 1 h and 30 s. During the EGM storage, an amplified signal from the storage capacitor can result in oversensing by the ventricular channel and inappropriate pacing inhibition. DISCUSSION: To rectify the issue, the ventricular lead sensitivity value was increased from 0.9 mV to 1.2 mV. No instances of inappropriate ventricular pacing inhibition were noted on follow-up. To our knowledge, this is a rare case of inappropriate ventricular pacing inhibition caused by a combination of PPM self-adjusting sensitivity algorithm and oversensing every 1 h and 30 s from an amplified storage capacitor. Physicians should be aware of this possible complication and differentiate it from device or lead malfunction.
No takes yet. Share an insight, caveat, or question.
Khan et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: