Key result
An 84-year-old woman experienced inappropriate high-rate pacing at 180 b.p.m. with maximal output due to a temporary pacemaker malfunction 11 days after insertion.
Case Report (n=1)
This case highlights a rare but severe runaway malfunction in a temporary pacemaker causing inappropriate high-rate pacing, likely due to oxidation-induced high resistance.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to rare temporary pacemaker runaway pacing; leaves open incidence, mechanisms, and prevention.
An 84-year-old woman presented with dizziness and dyspnoea due to high-degree atrioventricular block. After prompt insertion of single-chamber temporary pacemaker (Model 5348, VVI type, Serial No. PEP017919P, Medtronic, Minneapolis, MN, USA, yellow rectangles in Panel A), she underwent percutaneous coronary intervention (PCI) with stents at middle to the distal part of the right coronary artery owing to severe stenosis. The patient was closely monitored for unexpected arrhythmic events (temporary pacemaker backup rate of 50 pulses per minute, red arrow). However, 11 days after the PCI, she suddenly complained about marked dyspnoea and chest discomfort at 5 o'clock in the morning. The electrocardiogram, which was recorded during the event, demonstrated ventricular-paced rhythm at a rate of 180 b.p.m. and maximal output, which was never intended to set (Panel B). Fortunately, there was no further adverse event related to pacing rate out of the intended setting. She was safely discharged several days later after successful insertion of permanent pacemaker. According to the food and drug administration enforcement report in June 2014, this event seems to be caused by the development of high resistance on internal electrical connector contacts due to oxidation over time. To our knowledge, this is the first report showing inappropriate high-rate pacing with maximal output due to malfunction of temporary pacemaker. The full-length version of this report can be viewed at: http://www.escardio.
No takes yet. Share an insight, caveat, or question.
Park et al. (2016) conducted a case report in High-degree atrioventricular block (n=1). Single-chamber temporary pacemaker (Model 5348, VVI type) was evaluated on Inappropriate high-rate pacing with maximal output. An 84-year-old woman experienced inappropriate high-rate pacing at 180 b.p.m. with maximal output due to a temporary pacemaker malfunction 11 days after insertion.
Synapse has enriched 10 related papers. Consider them for comparative context: