In patients with sick sinus syndrome who develop AV conduction disease, backup ventricular pacing should be programmed to 'on (always)' rather than 'auto enable' during atrial ATP to prevent potentially catastrophic ventricular pauses.
Answers are grounded in this paper and cite their sources.
Suggests programming backup ventricular pacing to 'on (always)' during atrial ATP to avoid pauses; single case leaves open need for confirmation.
A 72-year-old woman was admitted to the hospital for episodes of lightheadedness and palpitations. She was diagnosed with SSS and underwent placement of a dual-chamber pacemaker (Medtronic Enrhythm) about a year ago. Her pacemaker interrogation is shown in Figure 1. A chest X-ray done at the time of this admission showed adequate placement of the pacing leads in the atrium and the right ventricular apex. In the hospital, telemetry strips were recorded during one of her episodes (Fig. 2B,C). Why is her device intermittently not pacing in the ventricle? Dual-chamber pacemaker at hospital admission. (A) P wave tracking. Panel (B) and panel (C) show telemetry strips recorded during two different episodes of presyncope. (D) Synch = synchronization interval; LRI = lower rate interval. Black arrows at PVCs. Atrial sensing and ventricular pacing are observed in Figure 2A, followed by a ventricular pause of 2.8 and 4 seconds in Figure 2B and Figure 2C, respectively. Atrial tachycardia of 220 ms cycle length is the underlying rhythm without ventricular conduction. There is no evidence of ventricular pacing spikes. An ECG monitor strip artifact must always be ruled out before reaching the diagnosis of pacemaker malfunction. The presence of several ECG strips recording with more than one lead showing the same finding with symptoms suggests that the ventricular pauses are real in this case. It is important and sometimes cumbersome to differentiate failure output with failure of capture1 because the electrical spikes may be difficult to see on current digital ECG recordings. However, here there are clear ventricular spikes at the beginning of the trace. Event markers electrograms are key, if available, to show absence of ventricular pacing, but these episodes were cleared out from the programmer hard drive before they could be reviewed. Careful analysis of the ECG strip (Fig. 2C) reveals the presence of pacing stimuli at about 200 ms, with no evidence of cardiac capture. Ventricular electrical output at 200 ms could be seen in a pacemaker as the result of two component failures of the pulse generator. “The runway pacemaker” is an unusual complication and requires urgent treatment. Intermittently rapid electrical spikes, as opposed to constant pacing in the “runaway pacemaker,” makes the latter unlikely. The presence of an atrial tachycardia prior to the pacemaker releasing rapid electrical stimuli is suggestive of antitachycardia pacing therapy being enabled and of a relationship between the ventricular standstill and the atrial therapies. The Enrhythm pacemaker detects an atrial tachycardia by examining the atrial rate and the relationship between atrial and ventricular event.2 Atrial ATP therapies are synchronized to the first atrial event after the ventricular event upon which the atrial arrhythmia is detected. The device can be programmed to deliver a sequence of up to three ATP therapies in the AT zone and up to three ATP therapies in the AT/AF if that zone is programmed. ATP atrial therapies can be programmed in the mode of burst or ramp. Atrial ATP cannot be delivered until the episode duration exceeds the program value from its detection (nominal 1 minute).2 Ventricular backup pacing in mode VVI is available during atrial ATP therapy delivery. It can be programmed off, on (always), and on (auto enable). The device is programmed on (auto enable) as default mode. When programmed on (always), backup pacing is delivered during every ATP therapy, if necessary. A backup ventricular output (VVI) of 6 V and 1.5 ms pulse width is then delivered at the programmed rate (nominal 60 bpm) or at the current pacing rate, whichever is faster, if there is no ventricular event sensed. When programmed on (auto enable), backup pacing is withheld unless all four ventricular events preceding ATP therapy were paced. In the event there is an intrinsic ventricular beat prior to the delivery of atrial therapies, a ventricular pause will occur. The pause in that case will be equal to the duration of an ATP sequence plus 500 ms (interval from AT/AF detection to atrial therapies) plus the post-ATP resynchronization time (interval from the last beat within ATP to the first paced beat = lower rate interval or the sensor rate, whichever is faster). Backup pacing was programmed on (auto enable) in this particular case. Unfortunately for the patient, there were frequent PVCs occurring within the last four ventricular events prior to the delivery of atrial therapies, and ventricular pacing was therefore inhibited. In Figure 2C, there is evidence of 17 pacemaker spikes (15A1 + 1S2 + 1S3). The tachycardia cycle length is about 220 ms; this resulted in a ventricular pause of about 4,000 ms. This ventricular pause is explained as follows: 500 ms (synchronization interval after AT/AF detection prior to delivering therapies) +ATP S1 (15 × 200 ms) +S2 (184 ms) +S3 (174 ms) +1,000 ms (lower rate limit). In this case, a PVC (arrow) occurred during the atrial synchronization interval (500 ms) and shortened the duration of ventricular standstill by 300 ms. The ventricular pause in Figure 2B is shorter than in Figure 2C, probably due to a different therapy of a shorter duration being delivered: for example, a ramp therapy. The other possibility is that the pacemaker sensor is driving the ventricular rate at a faster rate (780 ms) than the lower rate was set (1,000 ms), accounting for a shorter ventricular pause. The absence of the marker channels does not allow one to discern it. The worst scenario possible if backup pacing is not enabled and the device is programmed to DDD mode (in the case shown the device was set to VVI mode) would be a ventricular pause of about 15 sec.2 This patient developed advanced AV block within the last year. There is only about 1% annual incidence of developing AV conduction disease in patients with SSS, and its incidence is about 35% if there is an underlying bundle or bifascicular block.3 However, the potential of having significant ventricular pauses during ATP pacing could result in a catastrophic event. We recommend backup pacing to be programmed on (always) as default in order to avoid ventricular pauses in patients with SSS who develop A:V conduction disease.
No takes yet. Share an insight, caveat, or question.
ANTONIO NAVARRETE (2007) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: