Why the study?
Many patients with aortic regurgitation cannot undergo surgical replacement due to adverse surgical risk, and transcatheter valve implantation with most available prostheses has suboptimal success and outcomes.
The JenaValve Trilogy system offers a dedicated transcatheter option for high-surgical-risk patients with aortic regurgitation.
Answers are grounded in this paper and cite their sources.
Trilogy may offer an option for high-risk AR; leaves open need for prospective comparative data.
Aortic regurgitation (AR) is associated with morbidity and premature mortality. Surgical aortic valve replacement is not an option for many patients due to an adverse surgical risk profile, whilst transcatheter aortic valve implantation with most available prostheses has demonstrated suboptimal implantation success and outcomes. The JenaValve Trilogy™ system provides an attractive solution for such patients as it utilizes clips that directly attach onto the native valve leaflets to anchor. Initially designed for transapical delivery, the current transfemoral delivery system is under investigation in the United States and approved for aortic stenosis and regurgitation in Europe. We present an expert review on the technical aspects of the Trilogy system, provide a guide for implantation, discuss the available evidence for the technology and provide illustrative case examples.
No takes yet. Share an insight, caveat, or question.
Baumbach et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: