Key result
JenaValve Trilogy is anatomically feasible in ~54% of patients with isolated aortic regurgitation.
Why the study?
The proportion of all-comer patients with isolated aortic regurgitation unsuitable for surgery who have appropriate valve anatomy for the JenaValve Trilogy transcatheter heart valve was unknown. Furthermore, the specific reasons precluding placement of the prosthesis had not been systematically explored.
What proportion of patients with isolated symptomatic aortic regurgitation unsuitable for surgery have appropriate valve anatomy for transcatheter aortic valve implantation using the Trilogy Valve?
Cohort (n=50)
What proportion of patients with isolated symptomatic aortic regurgitation unsuitable for surgery have appropriate valve anatomy for transcatheter aortic valve implantation using the Trilogy Valve?
Approximately half (54%) of high-risk patients with isolated aortic regurgitation have suitable anatomy for TAVI with the Trilogy valve, with large annular dimensions being the primary reason for screening failure.
Answers are grounded in this paper and cite their sources.
Limits Trilogy TAVI to ~half of isolated AR patients; leaves open need for larger-annuli devices and prospective trials.
Background The introduction of a dedicated prosthesis for transcatheter treatment of patients with severe isolated aortic valve regurgitation (JenaValve Trilogy®) offers an interventional treatment option for patients unable to undergo surgical valve replacement. It is currently not known what proportion of all-comer patients with isolated aortic regurgitation unsuitable for surgery has appropriate valve anatomy for treatment with this transcatheter heart valve. Similarly, the specific reasons that preclude placement of the valve prosthesis have not been systematically explored. Methods An all-comer cohort of 107 consecutive patients with symptomatic aortic valve disease under consideration for transcatheter aortic valve implantation (TAVI) using the Trilogy Valve (JenaValve®) was included in this analysis. A detailed echocardiographic and computed tomographic (CT)-based assessment of aortic root anatomy and dimensions was performed. Mechanisms of aortic regurgitation, screening failure rate, and reasons for unsuitability were identified and compared between eligible and non-eligible patients. Results Out of 107 consecutive patients who underwent diagnostic workup prior to TAVI using the Trilogy Valve, 57 were excluded due to previous valve replacement, mixed aortic valve pathology including stenosis, or indications for TAVI using the Trilogy Valve other than isolated aortic regurgitation. Consequently, 50 patients were included (77±9 years, 22 females (44%), left ventricular ejection fraction (LVEF) 46±11%, vena contracta 6.5±1 mm, New York Heart Association (NYHA) class III in 50% and class II in 43%). Mechanisms of aortic regurgitation were as follows: 22% prolapse of the right coronary cusp, 20% prolapse of the non-coronary cusp, 16% annular dilatation, 34% degenerative disease, 4% bicuspid degenerative disease, 1 patient with prolapse of the left coronary cusp. Anatomy was suitable for implantation of the Trilogy Valve in 27 patients (54%), and unsuitable in 23 (46%). Reasons for screening failure included: large annular dimensions (61%), unfavourable aortic anatomy (22%), severe leaflet restriction (9%), bicuspid valve morphology (4%), and small annular dimensions (4%). Unsuitable patients had larger annular dimensions than suitable patients (mean diameter 27 vs 25.4 mm; area 588 vs 509 mm²; perimeter 87 vs 81 mm; all p<0.05). Other CT parameters such as the presence and degree of aortic valve calcification, aortic root angle, eccentricity index, and aortic dimensions were not significantly different between both groups. Conclusion Among patients with isolated aortic valve regurgitation at high risk for conventional surgery, approximately 50% were eligible for transcatheter treatment with the currently available models of the Trilogy heart valve. The most frequent reasons for non-eligibility include large annular dimensions or unfavourable anatomy of the aortic valve cusps.
No takes yet. Share an insight, caveat, or question.
Menhem et al. (2026) conducted a cohort in Isolated aortic regurgitation (n=50). JenaValve Trilogy TAVI eligibility assessment vs. Unsuitable valve anatomy was evaluated on Anatomical suitability for implantation of the Trilogy Valve (p=<0.05). Transcatheter treatment with the JenaValve Trilogy was anatomically feasible in 54% of patients with isolated aortic regurgitation, while 46% were unsuitable mainly due to large annular dimensions.