This case highlights the risk of very early S-ICD lead fracture and demonstrates that severe fibrous tissue encapsulation can occur rapidly, complicating surgical lead extraction.
Answers are grounded in this paper and cite their sources.
S-ICD lead fracture with fibrous encapsulation may complicate revisions; leaves open durability questions in young Brugada patients.
Six months after subcutaneous implantable cardioverter defibrillator (S-ICD) implantation a 26-year-old Brugada patient presented because of a beeping tone emitted by his device. Chest X-ray displayed two functionless transvenous shock leads and the S-ICD system with a lead fracture. During lead revision procedure, extensive preparation of the lead from unexpectedly firm surrounding fibrous tissue encapsulating the lead was necessary before it could be removed, and a new shock lead could be implanted. This is the first report of an S-ICD lead exchange due to very early lead fracture and unexpectedly severe fibrous tissue hampering surgical lead extraction.
No takes yet. Share an insight, caveat, or question.
Gutleben et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: