Why the study?
The S-ICD and its electrode were developed to avoid long-term complications of transvenous leads in the vasculature, but unexpected electrode fractures causing inappropriate shocks can still occur.
Clinicians should be aware of the potential for fatigue failure fracture of the S-ICD electrode, which can cause inappropriate shocks without preceding high-impedance alerts.
Answers are grounded in this paper and cite their sources.
S-ICD electrode fractures may cause inappropriate shocks without alerts; leaves open need for improved detection algorithms in larger cohorts.
The subcutaneous-implantable cardioverter-defibrillator (S-ICD) and its electrode were developed to avoid long-term complications of transvenous leads in the vasculature. We report a case of unexpected, inappropriate S-ICD shocks due to oversensing of high-amplitude, nonphysiologic, electrical noise artifacts that were not preceded by high-impedance alerts or sensing electrogram noise detections. Following explant, high-magnification X-ray imaging of the S-ICD electrode demonstrated partial fracture of the distal sensing conductor located near a short radius bend in the electrode at the electrode-header interface. Clinicians should be aware of a potential for fatigue failure fracture of the S-ICD electrode. Recommendations for systematic S-ICD follow-up and troubleshooting are discussed.