Key result
A 58-year-old female experienced inappropriate ICD shocks due to oversensing of high frequency noise caused by polyurethane lead failure, which was initially clinically silent at rest.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
ICD lead dysfunction can initially be clinically silent at rest and cause inappropriate shocks, highlighting the need for careful evaluation during long-term follow-up.
We describe here the case of a 58-year-old female patient who experienced inappropriate shocks from her implantable cardioverter-defibrillator (ICD). Stored electrograms from her ICD showed high frequency noise preceding the shock. Although the pacing threshold was normal and lead fracture was not found in chest X-rays, pacing lead impedance decreased to 480 omega. Moreover, such high frequency noise was observed by electrogram telemetry, but not by routine evaluation every 3 months. ICD lead dysfunction was suspected, so we elected to replace the ICD lead system. At the time of the operation, lead impedance was 410 omega and pacing threshold was the same as it was at the time of the ICD implantation, and no lead insulation disturbances were observed in the generator pocket. However, manipulation of the lead system produced high frequency noise reproducibly. Since some of the ICD lead dysfunction initially was clinically silent at rest, dysfunction was difficult to detect before serious problems occurred. Therefore, more careful evaluation of the ICD lead system is needed during long-term follow-up of ICD implants.
No takes yet. Share an insight, caveat, or question.
Washizuka et al. (2005) conducted a case report in Ventricular tachycardia and ICD lead failure (n=1). ICD lead failure was evaluated on Inappropriate shocks. A 58-year-old female experienced inappropriate ICD shocks due to oversensing of high frequency noise caused by polyurethane lead failure, which was initially clinically silent at rest.
Synapse has enriched 10 related papers. Consider them for comparative context: