Answers are grounded in this paper and cite their sources.
The pulmonary and systemic availability of budesonide after inhalation from a dry powder inhaler, Turbuhaler, and from a pressurized metered-dose inhaler (P-MDI) were compared in healthy volunteers. Two different methods were used to assess pulmonary availability: 1) calculated from the systemic availability corrected for an oral availability of 13% (n = 24); and 2) after blocking of gastrointestinal absorption by administration of a charcoal suspension (n = 13). An intravenous infusion of budesonide was used as a reference. The systemic availability of budesonide, calculated as a geometric mean and expressed as percentage of the metered dose, was 38% for Turbuhaler and 26% for P-MDI. The pulmonary availability, calculated using the first method, was 32% and 15% for Turbuhaler and P-MDI, respectively; and, using the second method, 32% and 18%, respectively. The results of the present study indicate that administration of budesonide via Turbuhaler gives rise to a lung deposition which is approximately twice that of a P-MDI, with less variability, but that systemic availability is only increased by approximately 50%. Thus, the present data suggest that by administrating budesonide via Turbuhaler, instead of a P-MDI, the same degree of asthma control can be achieved with a lower dose, which, in turn, reduces the risk of undesired systemic effects.
No takes yet. Share an insight, caveat, or question.
Thorsson et al. (1994) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: