Why the study?
Pulse oximeter photoplethysmography waveform measurements used to calculate peripheral capillary oxygen saturation are susceptible to motion noise from subject and sensor movements.
Does pre-filtering PPG signals with a heart-rate tuned comb filter improve the accuracy of SpO2 calculations during motion?
Does pre-filtering PPG signals with a heart-rate tuned comb filter improve the accuracy of SpO2 calculations during motion?
Using a heart-rate tuned comb filter to preprocess PPG signals improves the accuracy of pulse oximetry SpO2 measurements during motion artifacts.
Answers are grounded in this paper and cite their sources.
May warrant refined pulse oximetry algorithms during motion; leaves open large-scale clinical validation.
Calculation of peripheral capillary oxygen saturation [Formula: see text] levels in humans is often made with a pulse oximeter, using photoplethysmography (PPG) waveforms. However, measurements of PPG waveforms are susceptible to motion noise due to subject and sensor movements. In this study, we compare two [Formula: see text]-level calculation techniques, and measure the effect of pre-filtering by a heart-rate tuned comb peak filter on their performance. These techniques are: (1) "Red over Infrared," calculating the ratios of AC and DC components of the red and infrared PPG signals,[Formula: see text], followed by the use of a calibration curve to determine the [Formula: see text] level Webster (in: Design of pulse oximeters, CRC Press, Boca Raton, 1997); and (2) a motion-resistant algorithm which uses the Discrete Saturation Transform (DST) (Goldman in J Clin Monit Comput 16:475-83, 2000). The DST algorithm isolates individual "saturation components" in the optical pathway, which allows separation of components corresponding to the [Formula: see text] level from components corresponding to noise and interference, including motion artifacts. The comparison we provide here (employing the two techniques with and without pre-filtering) addresses two aspects: (1) accuracy of the [Formula: see text] calculations; and (2) computational complexity. We used both synthetic data and experimental data collected from human subjects. The human subjects were tested at rest and while exercising; while exercising, their measurements were subject to the impacts of motion. Our main conclusion is that if an uninterrupted high-quality heart rate measurement is available, then the "Red over Infrared" approach preceded by a heart-rate tuned comb filter provides the preferred trade-off between [Formula: see text]-level accuracy and computational complexity. A modest improvement in [Formula: see text] estimate accuracy at very low SNR environments may be achieved by switching to the pre-filtered DST-based algorithm (up to 6% improvement in [Formula: see text] level accuracy at -10 dB over unfiltered DST algorithm and the filtered "Red over Infrared" approach). However, this improvement comes at a significant computational cost.
No takes yet. Share an insight, caveat, or question.
Alkhoury et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: