Why the study?
Subcutaneous implantable cardioverter-defibrillator therapy is expanding rapidly, but data on the extraction procedure and subsequent patient management remain limited.
What are the procedural characteristics, management, and outcomes of S-ICD extraction in clinical practice?
What are the procedural characteristics, management, and outcomes of S-ICD extraction in clinical practice?
S-ICD extraction in clinical practice is safe, highly successful, and typically requires only simple manual traction, especially for systems with shorter dwell times.
Answers are grounded in this paper and cite their sources.
Supports safe, efficient S-ICD extraction in practice with short stays; leaves open need for randomized comparisons before guideline updates.
AIMS: Subcutaneous implantable cardioverter-defibrillator (S-ICD) therapy is expanding rapidly. However, there are few data on the S-ICD extraction procedure and subsequent patient management. The aim of this analysis was to describe the procedure, management, and outcome of S-ICD extractions in clinical practice. METHODS AND RESULTS: We enrolled consecutive patients who required complete S-ICD extraction at 66 Italian centres. From 2013 to 2022, 2718 patients undergoing de novo implantation of an S-ICD were enrolled. Of these, 71 required complete S-ICD system extraction (17 owing to infection). The S-ICD system was successfully extracted in all patients, and no complications were reported; the median procedure duration was 40 (25th-75th percentile: 20-55) min. Simple manual traction was sufficient to remove the lead in 59 (84%) patients, in whom lead-dwelling time was shorter [20 (9-32) months vs. 30 (22-41) months; P = 0.032]. Hospitalization time was short in the case of both non-infectious [2 (1-2) days] and infectious indications [3 (1-6) days]. In the case of infection, no patients required post-extraction intravenous antibiotics, the median duration of any antibiotic therapy was 10 (10-14) days, and the re-implantation was performed during the same procedure in 29% of cases. No complications arose over a median of 21 months. CONCLUSION: The S-ICD extraction was safe and easy to perform, with no complications. Simple traction of the lead was successful in most patients, but specific tools could be needed for systems implanted for a longer time. The peri- and post-procedural management of S-ICD extraction was free from complications and not burdensome for patients and healthcare system. CLINICAL TRIAL REGISTRATION: URL: http://clinicaltrials.gov/Identifier: NCT02275637.
No takes yet. Share an insight, caveat, or question.
Filippo et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: