Why the study?
The study was conducted to examine the rates of and indications for subcutaneous implantable cardioverter defibrillator (S-ICD) removal and extraction.
What are the rates, indications, and risk factors for S-ICD extraction in patients undergoing S-ICD implantation?
What are the rates, indications, and risk factors for S-ICD extraction in patients undergoing S-ICD implantation?
In a real-world cohort, S-ICD extraction occurred in only 5.9% of patients over 4.4 years, primarily due to pacing needs, infection, or oversensing, with smoking and higher BMI identified as independent risk factors.
Answers are grounded in this paper and cite their sources.
S-ICD extraction rates appear acceptably low in real-world use; observational data leaves open prospective evaluation of modifiable risks.
INTRODUCTION: The subcutaneous implantable cardioverter defibrillator (S-ICD) is an alternative to a transvenous ICD in patients who meet the criteria for ICD implantation without concurrent need for cardiac pacing. The objective of this study is to examine the rates of and indications for S-ICD removal and extraction. METHODS: A retrospective analysis of all patients who underwent S-ICD implantation between 2010 and 2022 at a single multihospital healthcare system was performed. The primary endpoint was S-ICD removal or extraction. Patient and device characteristics were abstracted from the electronic medical record. Univariate and multivariate analyses were completed to determine factors associated with S-ICD extraction. RESULTS: A total of 372 patients (69.5% male; 48.6 ± 14.4 years old) underwent S-ICD implantation during the study period. There were 22 (5.9%) patients (81.8% male; 52.1 ± 13.2 years old) who underwent S-ICD extraction over a median follow-up period of 4.4 [2.0-6.5] years. The median length of time between implantation and extraction was 39.6 [8.3-64.6] months. The most common indications for S-ICD extraction were the need for bradycardia pacing (frequency among extractions, 18.2%), infection (22.7%), and inappropriate shocks due to oversensing (22.7%). A smoking history and higher body mass index were independently associated with S-ICD extraction. CONCLUSIONS: The overall rate of S-ICD extraction over 4.4 [2.0-6.5] years was 5.9%, with the most common indications for extraction being the need for bradycardia pacing, infection, and inappropriate shocks due to oversensing. A smoking history and high body mass index are associated with increased rates of S-ICD extraction. With appropriate patient selection for the S-ICD, the need to remove the device after implantation is low.
No takes yet. Share an insight, caveat, or question.
Gangadharan et al. (2024) studied this question.