Wearable sensors enable continuous and non-invasive monitoring of biomarkers, facilitating home-based healthcare and reducing the reliance on costly laboratory tests. Here, we present a wearable electrochemical sensor based on laser-induced graphene (LIG) fabricated using polyimide and transferred onto a commercial adhesive bandage for the simultaneous detection of uric acid (UA) and ascorbic acid (AA) in sweat. The electrodes displayed a carbon-rich, homogeneous morphology with a hydrophilic surface, enhancing analyte interaction. The repeatability and reproducibility studies showed satisfactory relative standard deviations (RSD) values equal to 2.7 and 8.2%, respectively. Analytical performance for AA and UA was evaluated by differential pulse voltammetry (DPV) measurements in phosphate-buffered saline (PBS). In simultaneous detection, the sensor exhibited a linear response from 20 to 200 μmol l −1 (R 2 = 0.99) for AA and UA. The limit of detection (LOD) was 6.6 μmol l −1 and 6 μmol l −1 for AA and UA, respectively. The device was further applied in artificial sweat, exhibiting minimal interference (<10%) and recoveries of 96%–100% for AA and 90%–115% for UA (n = 3). These results demonstrate the potential of this low-cost (∼US$0.5/unit) flexible sensor as a practical platform for biofluid analysis.