Does cineradiography help identify paravalvular leaks in patients with suspected leaking cardiac prosthetic valves?
Does cineradiography help identify paravalvular leaks in patients with suspected leaking cardiac prosthetic valves?
Cineradiography can identify paravalvular leaks by measuring the angle of tilt of prosthetic valves, with specific thresholds for aortic (>6°) and mitral (>12°) valves.
Answers are grounded in this paper and cite their sources.
May support cineradiography for suspecting prosthetic valve leaks; hypothesis-generating and requires prospective validation before practice change.
Forty-nine patients whose clinical symptoms suggested a leaking cardiac prosthetic valve were studied by both cineradiography and angiocardiography. We found a wider range of tilting motion of the nonleaking mitral valve than has been previously shown. An angle of tilt greater than 6° for the prosthetic aortic valve strongly suggests separation of the prosthesis from its bed with a paravalvular leak, while the angle of tilt suggestive of a paravalvular leak for a prosthetic mitral valve is 12°. However, a tilt of less than these values does not entirely exclude the existence of a leak. We strongly suggest that routine cineradiography be done in each patient shortly after surgery, as a basis for comparison if the patient later develops a murmur or congestive heart failure.