Aims: An automated method for the measurement of blood tacrolimus on volumetric absorptive microsampling (VAMS) devices was developed. Materials & methods: VAMS devices prepared by the automated method were compared with those prepared by the existing manual method (n = 284; mean concentration: 8.0 μg/l; range: 0.6–18.1). Results: The performance of both methods was comparable. Passing–Bablok regression demonstrated an acceptable correlation (y = -0.449 + 1.06x). Bland-Altman analysis demonstrated acceptable agreement (mean bias: -0.007 μg/l; standard deviation: 1.536). Automation reduced operator touch time by 40 min (48-sample batch). Conclusion: Automated preparation of VAMS devices reduced touch time and improved process consistency, facilitating high-throughput testing and transformation of existing laboratory workflows. Automation did not improve precision for VAMS devices but did so for liquid blood samples.