Why the study?
Does EPI-accelerated PWV with improved spatial sampling improve the accuracy of aortic pulse wave velocity assessment compared to conventional PWV in patients with Marfan syndrome?
Does EPI-accelerated PWV with improved spatial sampling improve the accuracy of aortic pulse wave velocity assessment compared to conventional PWV in patients with Marfan syndrome?
Improving spatial sampling density with EPI-accelerated MRI yields more accurate aortic pulse wave velocity assessment in Marfan syndrome patients despite lower temporal resolution, while significantly reducing scan time.
Answers are grounded in this paper and cite their sources.
May support improved aortic PWV accuracy in Marfan syndrome; leaves open prospective validation before clinical adoption.
PURPOSE: To evaluate the effect of spatial (ie, number of sampling locations along the aorta) and temporal sampling density on aortic pulse wave velocity (PWV) assessment from velocity-encoded MRI in patients with Marfan syndrome (MFS). MATERIALS AND METHODS: Twenty-three MFS patients (12 men, mean age 36 ± 14 years) were included. Three PWV-methods were evaluated: 1) reference PWV(i.p.) from in-plane velocity-encoded MRI with dense temporal and spatial sampling; 2) conventional PWV(t.p.) from through-plane velocity-encoded MRI with dense temporal but sparse spatial sampling at three aortic locations; 3) EPI-accelerated PWV(t.p.) with sparse temporal but improved spatial sampling at five aortic locations with acceleration by echo-planar imaging (EPI). RESULTS: Despite inferior temporal resolution, EPI-accelerated PWV(t.p.) showed stronger correlation (r = 0.92 vs. r = 0.65, P = 0.03) with reference PWV(i.p.) in the total aorta, with less error (8% vs. 16%) and variation (11% vs. 27%) as compared to conventional PWV(t.p.) . In the aortic arch, correlation was comparable for both EPI-accelerated and conventional PWV(t.p.) with reference PWV(i.p.) (r = 0.66 vs. r = 0.67, P = 0.46), albeit 92% scan-time reduction by EPI-acceleration. CONCLUSION: Improving spatial sampling density by adding two acquisition planes along the aorta results in more accurate PWV assessment, even when temporal resolution decreases. For regional PWV assessment in the aortic arch, EPI-accelerated and conventional PWV assessment are comparably accurate. Scan-time reduction makes EPI-accelerated PWV assessment the preferred method of choice.
No takes yet. Share an insight, caveat, or question.
Kröner et al. (2012) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: