A novel recursive block-matching algorithm for 4D ultrasound left ventricle tracking demonstrates low segmentation errors of approximately 2.26-2.29 mm.
Answers are grounded in this paper and cite their sources.
Supports automated LV tracking in 4D ultrasound; leaves open prospective clinical validation before adoption.
We present a novel method for segmentation and tracking of the left ventricle (LV) in 4D ultrasound sequences using a combination of automatic segmentation at the end-diastolic frame and tracking using both a global optical flow-based tracker and local block matching. The core novelty of the proposed algorithm relies on the recursive formulation of the block-matching problem, which introduces temporal consistency on the patterns being tracked. The proposed method offers a competitive solution, with average segmentation errors of 2.29 and 2.26mm in the training (#=15) and testing (#=15) datasets respectively.
No takes yet. Share an insight, caveat, or question.
Barbosa et al. (2014) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: