Answers are grounded in this paper and cite their sources.
A transconductance driven-right-leg circuit improves high-frequency electromagnetic interference rejection by approximately 20 dB compared to classic designs, enhancing biopotential measurement quality.
Biopotential measurements are very sensitive to electromagnetic interference (EMI). EMI gets into the acquisition system by many ways, both as differential and common mode signals, driven-right-leg circuits (DRL) are widely used to reduce common mode interference. This paper reports an improvement on the classic DRL. The proposed circuit uses a transconductance amplifier to drive the patient's body. This configuration has some interesting properties, which provide an extended bandwidth for high-frequency EMI rejection (such as fluorescent lights interference). The improvement is around 20 dB for frequencies of few kilohertz and the circuit is easy to compensate for stability. A comparative analysis against a typical DRL is presented, the results obtained have been experimentally tested.
No takes yet. Share an insight, caveat, or question.
Spinelli et al. (1999) studied this question.