The proposed ECG amplifier design achieves high common-mode rejection and interference tolerance with low power consumption, potentially improving robust ECG acquisition.
Answers are grounded in this paper and cite their sources.
New ECG amp boosts TCMRR and CMI tolerance despite mismatch; leaves open clinical validation before practice adoption.
In this article, we present an electrocardiogram (ECG) amplifier that has a large total CMRR (TCMRR) regardless of contact impedance mismatch and a large tolerance to common-mode interference (CMI). To achieve these features, an adaptive TCMRR enhancing loop is implemented in parallel with a charge-pump-based common-mode suppressing loop (CMSL). It adjusts the CMI current through each contact impedance so that common-mode (CM) to differential-mode (DM) conversion is minimized. We also propose a fast settling technique for the adaptive loop so that contact impedance variation can be tracked fast enough to allow robust ECG acquisition. A prototype chip fabricated in 180-nm CMOS achieves TCMRR larger than 105 dB even when there is contact impedance mismatch of up to 30%. It also achieves tolerance to CMI of 18 VPP at 60 Hz and input-referred noise of 1.90μ Vᵣₘₛwhile consuming 43.3μ W.