Answers are grounded in this paper and cite their sources.
A regional supplementary comparison for the volume of liquid at 50 mL was conducted during October 2009 to June 2010 between the SIM members CENAM, INTI, INMETRO, INDECOPI and INTN. The transfer standard consisted of two 50 mL glass pycnometers, of the Gay Lussac type. CENAM acted as the pilot, collected the measurement results, analyzed the data and produced the comparison report. The median of all participants' results was used to calculate the regional comparison reference value because the result for one of the two pycnometers in one laboratory failed the chi-squared test at the 0.05 probability level. The measurements reported by the participants show an excellent overlap in four out of the five NMIs (−34×10−6 < Di < 29×10−6). The degree of equivalence obtained herein will be taken into account for the preparation of calibration and measurement capabilities claims from the participants. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the SIM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
No takes yet. Share an insight, caveat, or question.
Morales et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: