Why the study?
Can minute ventilation measured by a pacemaker sensor discriminate between sleep and waking hours in patients with a DDDR pacemaker?
Can minute ventilation measured by a pacemaker sensor discriminate between sleep and waking hours in patients with a DDDR pacemaker?
Minute ventilation measured by a pacemaker sensor can reliably discriminate between sleep and waking hours, offering a potential method to modulate backup pacing rates.
Answers are grounded in this paper and cite their sources.
May support nocturnal rate adaptation via pacemaker minute ventilation sensing; case report leaves open prospective validation before clinical use.
Special software allowing the memorization of 24-hour minute ventilation can be loaded into the memory of the Chorus RM, a DDDR pacemaker driven by minute ventilation. This feature was tested in the postimplant period in 13 patients. Measurements of minute ventilation, respiratory rate, and respiratory amplitude were analyzed according to prospectively defined diurnal and nocturnal time periods. Minute ventilation decreased by 39% (P < 0.001) from the diurnal to the nocturnal phase, while respiratory rate and amplitude decreased by 18% and 28%, respectively (P < 0.001 each). Thus, minute ventilation allowed discrimination between sleep and waking hours. This information could be utilized to modulate the backup rate of the pacemaker.
No takes yet. Share an insight, caveat, or question.
Bonnet et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: