Why the study?
Patients with epicardial implantable electronic devices are ineligible for MRI exams due to an elevated risk of RF heating.
Does modifying the trajectory of epicardial leads reduce MRI-induced RF heating in pediatric patients with CIEDs?
Does modifying the trajectory of epicardial leads reduce MRI-induced RF heating in pediatric patients with CIEDs?
Surgical modification of epicardial lead trajectory by looping excess length on the anterior surface of the heart substantially reduces MRI-induced RF heating at 1.5 T, potentially improving MRI safety for pediatric patients.
May permit MRI in select epicardial device patients; leaves open adoption pending prospective validation.
Purpose Infants and children with congenital heart defects, inherited arrhythmia syndromes, and congenital cardiac conduction disorders often receive epicardial implantable electronic devices. Unfortunately, once an epicardial device is implanted, the patient is no longer eligible to receive MRI exams due to an elevated risk of RF heating. Here we show that a simple modification in the trajectory of epicardial leads can substantially and reliably reduce RF heating during MRI at 1.5 T, with benefits extending to abandoned leads. Methods Electromagnetic simulations were performed to assess RF heating of two common epicardial lead trajectories exhibiting different degrees of coupling with MRI incident electric fields. Experiments in anthropomorphic phantoms implanted with commercial cardiac implantable electronic devices (CIEDs) confirmed the findings. Results Simulations of an epicardial lead with a trajectory where the excess length of the lead was looped and placed on the anterior surface of the heart showed a 9-fold reduction in 0.1g-averaged SAR compared to the lead with excess length looped on the inferior surface of the heart. Repeated experiments with a commercial epicardial device confirmed the results, showing a 16-fold reduction in the average temperature rise for fully implanted systems with leads following low-SAR trajectories, and a 20-fold reduction in RF heating on an abandoned lead. Conclusion Surgical modification of epicardial lead trajectory can substantially reduce RF heating at 1.5 T, with benefits extending to abandoned leads.
No takes yet. Share an insight, caveat, or question.
Jiang et al. (2022) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: