Key result
Adapt aberration correction improves echo image coherence by ~62%, earning clinician preference in 97% of loops.
Why the study?
Does the Adapt receive-only aberration correction algorithm improve 2D transthoracic echocardiographic image quality and clinician preference compared with standard beamforming in adult cardiology patients?
Does the Adapt receive-only aberration correction algorithm improve 2D transthoracic echocardiographic image quality and clinician preference compared with standard beamforming in adult cardiology patients?
Receive-only aberration correction systematically enhances 2D echocardiographic image sharpness, endocardial border definition, and tissue-to-cavity contrast, earning a 97% preference rate among evaluating cardiologists.
Answers are grounded in this paper and cite their sources.
Should not yet change echocardiography practice; hypothesis-generating for randomized trials on diagnostic impact.
Background: An aberration correction algorithm has been implemented and demonstrated in an echocardiographic clinical trial using two-dimensional (2D) imaging. The method estimates and compensates arrival time errors between different sub-aperture processor (SAP) signals in a matrix array probe. Methods: Five standard views of channel data cineloops were recorded from 22 patients (11 male and 11 female) resulting in a total of 116 cineloops. The channel data were processed with and without the aberration correction algorithm, allowing for side-by-side comparison of images processed from the same channel data cineloops. Results: The aberration correction algorithm improved image quality, as quantified by a coherence metric, in all 7,380 processed frames. In a blinded and left-right-randomized side-by-side evaluation, four cardiologists (two experienced and two in training) preferred the aberration corrected cineloops in 97% of the cases. The clinicians reported that the corrected cineloops appeared sharper with better contrast and less noise. Many structures like valve leaflets, chordae, endocardium, and endocardial borders appeared narrower and more clearly defined in the aberration corrected images. An important finding is that aberration correction improves contrast between the endocardium and ventricle cavities for every processed image. The gain difference was confirmed by the cardiologists in their feedback and quantified with a median global gain difference estimate between the aberration-corrected and non-corrected images of 1.2 dB. Conclusions: The study shows the potential value of aberration correction in clinical echocardiography. Systematic improvement of images acquired with state-of-art equipment was observed both with quantitative metrics of image quality and clinician preference.
No takes yet. Share an insight, caveat, or question.
Måsøy et al. (2023) studied Clinical indication for transthoracic echocardiography (TTE) (n=22). Adapt aberration correction algorithm (receive-only arrival time error compensation in cSound software beamforming) vs. Standard cSound image processing without aberration correction on the same recorded channel data cineloops was evaluated on Image quality improvement assessed by clinician preference in side-by-side paired cineloop evaluation and Global Image Coherence (GIC) (Median average GIC increase 62%; Clinician preference 97.4% vs 2.4%, p=<0.05). The Adapt aberration correction algorithm increased global image coherence across all 7,380 frames and was preferred by clinicians in 97% of 116 cineloops.
Synapse has enriched 10 related papers. Consider them for comparative context: