Key result
4D PC MRI quantifies LV kinetic energy with low bias between 1.5 T and 3 T.
Why the study?
The kinetic energy content of blood in the left ventricle during the cardiac cycle had previously not been described, prompting the development of a calculation method using 4D phase-contrast MRI.
Is it feasible to quantify left ventricular kinetic energy during the cardiac cycle using 4D phase-contrast MRI across 1.5 T and 3 T scanners in healthy volunteers?
Observational (n=6)
No
Is it feasible to quantify left ventricular kinetic energy during the cardiac cycle using 4D phase-contrast MRI across 1.5 T and 3 T scanners in healthy volunteers?
This feasibility study demonstrates that 4D phase-contrast MRI can non-invasively quantify left ventricular blood kinetic energy across the cardiac cycle with low variability between 1.5 T and 3 T scanners.
Answers are grounded in this paper and cite their sources.
Supports scanner-agnostic LV kinetic energy quantification in volunteers; leaves open validation and utility in patients.
Methods Six healthy volunteers age 32 ± 11 (mean ± SD), 4 men, were imaged in a 1.5 and 3 T Philips scanner, i.e. two scans per subject in one day. A 4D PC TFE-sequence with retrospective ECG-triggering and respiratory navigator, 3 mm isotropic voxels and a temporal resolution of 25 ms interpolated to 40 phases was used to quantify intraventricular flow. Typical sequence parameters: 1.5 T SENSE factor 2, TE/TR/flip (3.7/6.3 ms/8°). 3 T, SENSE factor 2, TE/TR/flip (3.7/6.2 ms/8°). Cine images from a SSFP sequence (TE/TR/flip, 1.5/2.9 ms/60°) with spatial resolution 1.48 × 1.48 mm, slice thickness 8 mm and a temporal resolution of 30 heart phases, covering the LV in short axis view was used to delineate the endocardial contours. 4D-flow data was post-processed using an in-house developed software including linear background phase correction. PC images were reconstructed from the 4Ddata in the short axis view and the endocardial contours were imported and manually corrected when needed. The kinetic energy was calculated as the squared velocity times the mass divided by two for all voxels within the LV for each time point. LV energy values are presented as mean ± SEM and results from the 1.5 T and 3 T measurements were compared according to Bland-Altman analysis (mean ± SD). from 13th Annual SCMR Scientific Sessions Phoenix, AZ, USA. 21-24 January 2010
No takes yet. Share an insight, caveat, or question.
Wåxnäs et al. (2010) conducted an observational in Healthy volunteers (n=6). 4D velocity encoded phase contrast (PC) MRI (3 T and 1.5 T) vs. Comparison between 1.5 T and 3 T MRI scanners was evaluated on Left ventricular kinetic energy (systolic, early diastolic, and late diastolic peaks). 4D PC MRI quantified three LV kinetic energy peaks with low bias between 1.5 T and 3 T scanners.