Does the Lotus Valve prosthesis improve hemodynamics and symptoms in a high-risk patient with severe aortic stenosis?
Does the Lotus Valve prosthesis improve hemodynamics and symptoms in a high-risk patient with severe aortic stenosis?
Answers are grounded in this paper and cite their sources.
First-in-man experience demonstrates the feasibility and short-term efficacy of percutaneous implantation of the repositionable Lotus Valve in a high-risk patient with severe aortic stenosis.
OBJECTIVES AND BACKGROUND: Percutaneous aortic valve replacement is a new less-invasive alternative for high-risk surgical candidates with aortic stenosis. However, the clinical experience is still limited, and the currently available 'first-generation devices' revealed technical shortcomings, such as lack of repositionability and presence of paravalvular leakages. We report the first-in-man experience with the new self-expanding Lotus Valve prosthesis composed of a nitinol frame with implemented bovine pericardial leaflets which is designed to address these issues, being repositionable and covered by a flexible membrane to seal paravalvular gaps. We implanted this prosthesis in a 93-year old patient presenting with severe symptomatic aortic stenosis (valve area: 0.6 cm(2)). Surgical valve replacement had been declined due to comorbidities. METHODS AND RESULTS: We used a retrograde approach for insertion of the 21-French Lotus catheter loaded with the valve prosthesis via surgical cut-down to the external iliac artery. Positioning of the valve was guided by transesophageal echo and supra-aortic angiograms. The prosthesis was successfully inserted and deployed within the calcified native valve. Echocardiography immediately after device deployment showed a significant reduction of the transaortic mean pressure gradient (32 to 9 mmHg; final valve area 1.7 cm(2)) without evidence of residual aortic regurgitation. The postprocedural clinical status improved from NYHA-IV to NYHA-II. These results remained unchanged up to the 3 month follow-up. CONCLUSIONS: Successful percutaneous aortic valve replacement can be performed using the new self-expanding and repositionable Lotus valve for treatment of high-risk patients with aortic valve stenosis. Further studies are mandatory to assess device safety and efficacy in larger patient populations.
No takes yet. Share an insight, caveat, or question.
Buellesfeld et al. (2008) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: