Why the study?
Does single-coil defibrillation (SVC coil off) provide a satisfactory defibrillation safety margin compared to dual-coil defibrillation in patients with hypertrophic cardiomyopathy?
Does single-coil defibrillation (SVC coil off) provide a satisfactory defibrillation safety margin compared to dual-coil defibrillation in patients with hypertrophic cardiomyopathy?
Single-coil ICD leads provide a satisfactory defibrillation safety margin for left-sided implantation in patients with hypertrophic cardiomyopathy, despite higher shock impedance and pulse width compared to dual-coil leads.
Answers are grounded in this paper and cite their sources.
Single-coil ICD leads may be feasible in HCM; leaves open need for larger studies on thresholds and outcomes.
BACKGROUND: Single-coil defibrillator leads have gained favor because of their potential ease of extraction. However, a high defibrillation threshold remains a concern in patients with hypertrophic cardiomyopathy (HCM), and in many cases, dual-coil leads have been used for this patient group. There is little data on using single-coil leads for HCM patients. METHODS AND RESULTS: We evaluated 20 patients with HCM who received an implantable cardioverter-defibrillator (ICD) on the left side in combination with a dual-coil lead. Two sets of defibrillation tests were performed in each patient, one with the superior vena cava (SVC) coil "on" and one with the SVC coil "off". ICDs were programmed to deliver 25 joules (J) for the first attempt followed by maximum energy (35 J or 40 J). Shock impedance and shock pulse width at 25 J in each setting as well as the results of the shock were analyzed. All 25-J shocks in both settings successfully terminated ventricular fibrillation. However, shock impedance and pulse width increased substantially with the SVC coil programmed "off" compared with "on" (66.4±6.1 ohm and 14.0±1.3 ms "off" vs. 41.9±5.0 ohm and 9.3±0.8 ms "on", P<0.0001 respectively). CONCLUSIONS: Biphasic 25-J shocks with the SVC coil 'off' successfully terminated ventricular fibrillation in HCM patients, indicating a satisfactory safety margin for 35-J devices. Single-coil leads appear appropriate for left-sided implantation in this patient group. (Circ J 2016; 80: 2199-2203).
No takes yet. Share an insight, caveat, or question.
Okamura et al. (2016) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: