Key result
A dual coil defibrillation lead system showed no significant advantage over a single coil system for defibrillation thresholds at implantation.
Why the study?
Does a dual coil defibrillation lead system improve defibrillation thresholds and lead parameters compared to a single coil system in patients with ventricular arrhythmias undergoing ICD therapy?
RCT (n=76)
randomized
Yes
Does a dual coil defibrillation lead system improve defibrillation thresholds and lead parameters compared to a single coil system in patients with ventricular arrhythmias undergoing ICD therapy?
Dual coil defibrillation leads offer no significant advantage over single coil leads regarding defibrillation thresholds or lead parameters during ICD implantation.
Answers are grounded in this paper and cite their sources.
Single-coil leads suffice for defibrillation thresholds; confirms no advantage of dual-coil systems in this RCT.
ICD implantation is standard therapy for malignant ventricular arrhythmias. The advantage of dual and single coil defibrillator leads in the successful conversion of arrhythmias is unclear. This study compared the effectiveness of dual versus single coil defibrillation leads. The study was a prospective, multicenter, randomized study comparing a dual with a single coil defibrillation system as part of an ICD using an active pectoral electrode. Seventy-six patients (64 men, 12 women; age 61 +/- 11 years) were implanted with a dual (group 1, n = 38) or single coil lead system (group 2,n = 38). The patients represented a typical ICD cohort: 60% presented with ischemic cardiomyopathy as their primary cardiac disease, the mean left ventricular ejection fraction was 0.406 +/- 0.158. The primary tachyarrhythmia was monomorphic ventricular tachyarrhythmia in 52.6% patients and ventricular fibrillation in 38.4%. There was no significant difference in terms of P and R wave amplitudes, pacing thresholds, and lead impedance at implantation and follow-up in the two groups. There was similarly no difference in terms of defibrillation thresholds (DFT) at implantation. Patients in group 1 had an average DFT of 10.2 +/- 5.2 J compared to 10.3 +/- 4.1 J in Group 2, P = NS. This study demonstrates no significant advantage of a dual coil lead system over a single coil system in terms of lead values and defibrillation thresholds. This may have important bearing on the choice of lead systems when implanting ICDs.
No takes yet. Share an insight, caveat, or question.
Rinaldi et al. (2003) conducted an RCT in Ventricular arrhythmias (n=76). Dual coil defibrillation lead system vs. Single coil defibrillation lead system was evaluated on Defibrillation thresholds (DFT) at implantation (p=NS). A dual coil defibrillation lead system showed no significant advantage over a single coil system for defibrillation thresholds at implantation.