Key result
Single-coil ICDs reduce all-cause mortality ~9% vs dual-coil despite worse defibrillation thresholds.
Why the study?
Data comparing the efficacy and extraction-related outcomes of single-coil versus dual-coil ICDs remained conflicting.
Does single-coil ICD compared to dual-coil ICD improve efficacy and extraction-related outcomes in patients requiring ICDs?
Meta-Analysis
Does single-coil ICD compared to dual-coil ICD improve efficacy and extraction-related outcomes in patients requiring ICDs?
While single-coil ICDs have shorter implant times and lower all-cause mortality, dual-coil ICDs offer superior defibrillation thresholds with no significant difference in extraction-related complications.
Answers are grounded in this paper and cite their sources.
Single-coil ICDs reduce mortality despite inferior thresholds; supports preferring them over dual-coil devices in routine practice.
BACKGROUND: Implantable cardioverter defibrillators (ICD) are battery-operated devices used to manage irregular heart rhythms and deliver therapeutic shocks to the heart. This updated systematic review and meta-analysis compares the efficacy and extraction-related outcomes of single-coil versus dual-coil ICDs in view of conflicting data. METHODS: Several databases, including PubMed, Cochrane Library, and Google Scholar, were comprehensively explored dating from inception to April 1, 2024. Data were compared using odds ratio (OR), hazard ratio (HR), and mean differences (MD). A value of p < 0.05 indicated statistical significance. RESULTS: Ultimately, 28 studies were included in this quantitative synthesis. Defibrillation threshold (DFT) indicated statistical superiority in the dual-coil cohort (MD: 0.58; 95% confidence interval [CI]: 0.07-1.09; p = 0.03). In addition, all-cause mortality was significantly elevated in the dual-coil cohort (HR: 0.91; 95% CI: 0.87-0.97; p = 0.001). Furthermore, implant time was significantly lower in the single-coil group (MD: -7.44; 95% CI: -13.44 to -1.43; p = 0.02). Other outcomes, including first shock efficacy, cardiac mortality, post-extraction major complications, post-extraction procedural success, and post-extraction mortality, did not demonstrate any significant statistical differences. CONCLUSION: In conclusion, despite the desirable safety profile of single-coil ICDs, the use of dual-coil ICDs continues to hold merit due to their superior efficacy and advanced sensing capabilities, especially in complex cases. In addition, the perceived risk of a greater adverse profile in dual-coil lead extraction can be refuted by preliminary aggregate results generated within this meta-analysis. However, further robust studies are warranted before arriving at a valid conclusion.
No takes yet. Share an insight, caveat, or question.
Tahir et al. (2025) conducted a meta-analysis in Irregular heart rhythms. Single-coil implantable cardioverter defibrillators vs. Dual-coil implantable cardioverter defibrillators was evaluated on Defibrillation threshold (DFT) (MD 0.58, 95% CI 0.07-1.09, p=0.03). Dual-coil ICDs showed superior defibrillation thresholds but higher all-cause mortality compared to single-coil ICDs.
Synapse has enriched 10 related papers. Consider them for comparative context: