Generic ICD programming recommendations can interact with manufacturer-specific features to withhold life-saving therapy for ventricular fibrillation, leading to sudden death.
Answers are grounded in this paper and cite their sources.
Generic thresholds risk untreated VF in select ICDs; leaves open manufacturer-specific programming validation in trials.
BACKGROUND: In clinical trials, manufacturer-specific, strategic programming of implantable cardioverter-defibrillators (ICDs), including faster detection rates, reduces unnecessary therapy but permits therapy for ventricular tachycardia/ventricular fibrillation (VF). Present consensus recommends a generic rate threshold between 185 and 200 beats per minute, which exceeds the rate tested in clinical trials for some manufacturers. In a case series, we sought to determine the relationship between programmed parameters and failure of modern ICDs to treat VF. METHODS AND RESULTS: We reviewed cases in which normally functioning ICDs failed to deliver timely therapy for VF from April 2015 to January 2017 at 4 institutions. Of 10 ambulatory patients, 5 died from untreated VF, 4 had cardiac arrests requiring external shocks, and 1 was rescued by a delayed ICD shock. VF did not satisfy programmed detection criteria in 9 patients (90%). Seven of these patients had slowest detection rates that were consistent with generic recommendations but not tested in a peer-reviewed trial for their manufacturer's ICDs. Manufacturer-specific factors interacted with fast detection rates to withhold therapy, including strict VF episode termination rules, enhancements to minimize T-wave oversensing, and features that restrict therapy to regular rhythms in ventricular tachycardia zones. Untreated VF despite recommended programming accounted for 56% of sudden deaths and 11% of all deaths during the study period. CONCLUSIONS: Complex and unanticipated interactions between manufacturer-specific features and generic programming can prevent therapy for VF. More data are needed to assess the risks and benefits of translating evidence-based detection parameters from one manufacturer to another.
No takes yet. Share an insight, caveat, or question.
Thøgersen et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: