Why the study?
Does a novel discrimination algorithm reduce inappropriate therapy due to T-wave oversensing in the S-ICD system?
Population
Database of recorded episodes from subcutaneous ICD systems, including a development set and an independent…
Comparison
A novel discrimination algorithm for the S-ICD… vs The S-ICD system without the new algorithm.
Design
Other
Authors
Loading...
May reduce inappropriate S-ICD therapies; leaves open prospective validation before practice change.
Does a novel discrimination algorithm reduce inappropriate therapy due to T-wave oversensing in the S-ICD system?
A novel discrimination algorithm for the S-ICD system significantly reduces inappropriate charges due to T-wave oversensing without compromising appropriate therapy for ventricular arrhythmias.
Brisben et al. (2015) studied this question.