Key result
S-ICD inappropriate shocks affect ~8% of patients, linked to taller height and first-generation devices.
Why the study?
Inappropriate shocks remain a challenge for patients and physicians after implantation of the subcutaneous implantable cardioverter-defibrillator (S-ICD).
What are the incidence, mechanisms, and predictors of inappropriate shocks in patients with a subcutaneous defibrillator?
Cohort (n=239)
What are the incidence, mechanisms, and predictors of inappropriate shocks in patients with a subcutaneous defibrillator?
Answers are grounded in this paper and cite their sources.
Inappropriate shocks in S-ICD patients occur in about 8% of patients, are most often caused by myopotentials on the primary sensing vector, and are associated with increased rehospitalization.
AIMS: Inappropriate shocks (IAS) remain a challenge for patients and physicians after implantation of the subcutaneous implantable cardioverter-defibrillator (S-ICD). The aims were to assess and characterize different patterns of IAS. METHODS AND RESULTS: Two hundred and thirty-nine patients were implanted with an S-ICD between 2010 and 2018 for primary and secondary prevention. Follow-up data of at least 6 months were analysed. During a mean follow-up of 34.9 ± 16.0 months, a total of 73 shocks occurred in 38 patients (6%). Forty-three (59%) shocks were considered appropriate due to ventricular tachycardia/ventricular fibrillation, while 30 (41%) were inappropriate and occurred in 19 patients (8%). Myopotentials/noise was the most frequent cause of inappropriate shocks (n = 8), followed by T-wave oversensing (n = 6) and undersensing of the QRS, resulting in adaptation of the automatic gain control and inappropriate shock (n = 5). Seventy-four percent of all IAS occurred on the primary vector, while no IAS occurred on the alternate vector. In seven of eight patients (88%), IAS related to myopotentials have occurred on the primary sensing vector. Multivariate analysis identified taller patients, primary sensing vector and first-generation S-ICD device as predictors for IAS. SMART pass effectively reduced the occurrence of IAS in the second-generation S-ICD system. CONCLUSION: Inappropriate therapies are less frequently observed on the alternate vector. The primary vector seems to be unfavourable with regard to oversensing caused by myopotentials. Inappropriate shocks were associated with an increased rate of rehospitalization but not mortality. These observations have implications for the prevention of inappropriate S-ICD shocks.
No takes yet. Share an insight, caveat, or question.
Rudic et al. (2020) conducted a cohort in S-ICD implantation (n=239). Subcutaneous implantable cardioverter-defibrillator (S-ICD) was evaluated on Inappropriate shocks (IAS). Among patients with an S-ICD, inappropriate shocks occurred in 8% over 34.9 months, with taller patients, primary sensing vector, and first-generation devices identified as predictors.