Key result
S-ICDs linked to ~113% higher risk of inappropriate shocks from noise oversensing vs EV-ICDs.
Why the study?
Extracardiac sensing electrode locations in S-ICDs and EV-ICDs may increase susceptibility to oversensing and inappropriate shocks, but how the causes of inappropriate shocks compare between devices was unknown.
Do the causes of inappropriate shocks differ between subcutaneous and extravascular implantable cardioverter-defibrillators?
Observational (n=1,968)
Do the causes of inappropriate shocks differ between subcutaneous and extravascular implantable cardioverter-defibrillators?
The causes of inappropriate shocks differ significantly between S-ICDs and EV-ICDs, with S-ICDs more prone to T-wave oversensing and EV-ICDs more prone to P-wave oversensing and misinterpreting supraventricular arrhythmias.
Answers are grounded in this paper and cite their sources.
May increase inappropriate shocks from noise oversensing with S-ICDs; leaves open optimal device selection pending prospective data.
Introduction Subcutaneous implantable cardioverter defibrillators (S-ICDs) and extravascular-implantable cardioverter-defibrillators (EV-ICDs) have emerged as alternative defibrillation options to avoid the complications of transvenous leads. However, their sensing electrodes' extracardiac locations may increase their susceptibility to oversensing which may cause painful, and potentially arrhythmogenic, inappropriate shocks (IAS). Objective The objective of this study was to compare the causes of IAS between the S-ICD and EV-ICD. Methods The FDA’s Manufacturer and User Facility Device Experience (MAUDE) database was queried for IAS reports submitted by the two manufacturers from January 2024 to September 2025. Outcomes included 1) oversensing noise, T-waves oversensing (TWOS), P-waves oversensing (PWOS), and electromagnetic interference (EMI), and 2) misinterpreting supraventricular tachycardia (SVT), atrial fibrillation (AF), and sinus tachycardia (ST) as ventricular tachycardia (VT) or ventricular fibrillation (VF). Results The search found 1,900 S-ICD and 68 EV-ICD IAS reports. Oversensing of unspecified noise and TWOS caused significantly more IAS by S-ICDs vs EV-ICDs (RR = 2.13, CI: 1.24-3.68, p<0.01; RR = 8.88, CI: 2.26-34.84, p<0.01, respectively), while IAS due to misinterpreting SVT, AF, and ST as VT/VF (RR = 0.32, CI: 0.24-0.44, p<0.01), PWOS (RR = 0.1, CI: 0.05-0.21, p<0.01), and EMI (RR = 0.15, CI: 0.09-0.25, p<0.01) were more common with EV-ICDs than SC-ICDs (Table 1, Figure 1). Oversensing of myopotential noise was not different between the two groups. Only with S-ICD, sense B-noise, air entrapment, and intrinsic morphology changes were causes of IAS (Table 1). In 8 S-ICD and 1 EV-ICD reports, there was more than one cause of IAS. Conclusion The causes of IAS are significantly different between S-ICDs and EV-ICDs. The incidence and reasons for these major differences as well as the long-term clinical consequences of these differences require further investigation. If confirmed, this may assist in individualizing device selection to patient-specific risk profiles and encourage targeted follow-up strategies.
No takes yet. Share an insight, caveat, or question.
Abdelsayed et al. (2026) conducted an observational in Inappropriate shocks in implantable cardioverter-defibrillators (n=1,968). Subcutaneous implantable cardioverter-defibrillators (S-ICDs) vs. Extravascular-implantable cardioverter-defibrillators (EV-ICDs) was evaluated on Oversensing of unspecified noise as a cause of inappropriate shocks (RR 2.13, 95% CI 1.24-3.68, p=<0.01). S-ICDs were associated with more inappropriate shocks due to unspecified noise oversensing and T-wave oversensing compared to EV-ICDs.