Key result
Reprogramming S-ICD sensing vector to secondary eliminates P- and T-wave oversensing, preventing inappropriate shocks.
Why the study?
Does reprogramming the sensing vector resolve inappropriate S-ICD shocks caused by combined P- and T-wave oversensing in a patient with HFrEF in sinus rhythm?
Case Report (n=1)
No
Does reprogramming the sensing vector resolve inappropriate S-ICD shocks caused by combined P- and T-wave oversensing in a patient with HFrEF in sinus rhythm?
This case demonstrates that combined P- and T-wave oversensing causing inappropriate S-ICD shocks can occur in patients in sinus rhythm without hypertrophic obstructive cardiomyopathy, particularly in the setting of low-amplitude QRS complexes and atrial enlargement, and can be resolved by modifying the sensing vector.
Answers are grounded in this paper and cite their sources.
Sensing vector reprogramming may avert inappropriate S-ICD shocks from combined oversensing; leaves open generalizability beyond single cases.
Subcutaneous implantable cardiac defibrillators (S-ICD) are an emerging technology that, in appropriately selected individuals, can be an acceptable alternative to transvenous ICDs (TV-ICD), which can lead to reduced complications associated with lead placement and potential improved durability. S-ICDs have comparable inappropriate discharge rates when compared to TV-ICDs, the cause of which in commonly supraventricular tachycardia (SVT) or T-wave or extracardiac oversensing.1 While effective and generally safe, TV-ICDs do require placement of TV leads, which are the source of many of the complications.
No takes yet. Share an insight, caveat, or question.
Ochman et al. (2020) conducted a case report in Heart failure with reduced ejection fraction (LVEF 20%) and subcutaneous implantable cardiac defibrillator (S-ICD) experiencing inappropriate shocks due to P- and T-wave oversensing (n=1). Modification of S-ICD sensing vector from primary to secondary sensing vector and adjustment of conditional/shock zone thresholds (from 180/240 bpm to 200/250 bpm) was evaluated on Resolution of P- and T-wave oversensing and recurrent inappropriate S-ICD discharges. Modification of the S-ICD sensing vector from primary to secondary resolved simultaneous P- and T-wave oversensing and prevented further inappropriate shocks in a 55-year-old man in sinus rhythm.