Key result
A patient with a subcutaneous implantable cardioverter-defibrillator experienced inappropriate shocks attributed to the rare triple counting of P-, R-, and T-waves while in sinus rhythm.
Why the study?
Inappropriate S-ICD shocks due to triple counting of P-, R-, and T-waves are an unusual occurrence with only a limited number of documented case reports.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
This case highlights a rare occurrence of inappropriate S-ICD shocks caused by triple counting of P-, R-, and T-waves in sinus rhythm, emphasizing the need for careful vector screening and anatomical consideration during implantation.
Sudden cardiac death (SCD) caused by ventricular tachyarrhythmias is a significant contributor to cardiovascular deaths worldwide. Implantable cardioverter-defibrillators (ICDs) have shown efficacy in preventing and reducing mortality from SCD, but traditional transvenous ICDs have inherent challenges and drawbacks, such as lead fractures, lead-associated endocarditis, and lead failure. To address these issues, subcutaneous ICDs (S-ICDs) have been developed. S-ICDs lack pacing capacity but are a valid alternative for patients at high risk for infection or with difficult venous access. Pre-implantation screening can help prevent inappropriate device shocks. We present a case in which a patient received inappropriate S-ICD therapy, which was attributed to the triple counting of P-, R-, and T-waves in a patient with sinus rhythm. This is an unusual occurrence, and, to the best of our knowledge, there are only a limited number of case reports documenting inappropriate shocks due to the oversensing of P-waves and T-waves.
No takes yet. Share an insight, caveat, or question.
Ghanta et al. (2023) conducted a case report in Non-ischemic cardiomyopathy (n=1). Subcutaneous implantable cardioverter-defibrillator (S-ICD) was evaluated on Inappropriate device shock. A patient with a subcutaneous implantable cardioverter-defibrillator experienced inappropriate shocks attributed to the rare triple counting of P-, R-, and T-waves while in sinus rhythm.