Key result
An initial appropriate ICD shock induced persistent cellular repolarization abnormalities, resulting in T-wave oversensing that triggered 84 inappropriate shocks and rapid battery depletion.
Population
1 67-year-old female with valvular cardiomyopathy and permanent atrial fibrillation, implanted with a…
Design
Case_report
Authors
Loading...
Alerts clinicians to post-shock T-wave oversensing risk; leaves open prevalence and mechanisms in larger cohorts.
Case Report (n=1)
No
An appropriate ICD shock may induce persistent cellular repolarization abnormalities, leading to T-wave oversensing, multiple inappropriate shocks, and rapid battery depletion.
Campana (2014) conducted a case report in T-wave oversensing and inappropriate ICD shocks (n=1). Implantable Cardioverter Defibrillator (ICD) was evaluated on Inappropriate shocks and battery depletion. An initial appropriate ICD shock induced persistent cellular repolarization abnormalities, resulting in T-wave oversensing that triggered 84 inappropriate shocks and rapid battery depletion.