Key result
An initial appropriate ICD shock induced persistent cellular repolarization abnormalities, resulting in T-wave oversensing that triggered 84 inappropriate shocks and rapid battery depletion.
Case Report (n=1)
No
An appropriate ICD shock may induce persistent cellular repolarization abnormalities, leading to T-wave oversensing, multiple inappropriate shocks, and rapid battery depletion.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to post-shock T-wave oversensing risk; leaves open prevalence and mechanisms in larger cohorts.
Implantable cardioverter-defibrillators (ICDs) have a considerable impact on the therapy of patients with threatening ventricular arrhythmias, significantly reducing the risk of sudden death. Even though newest devices have achieved a remarkable reliability, the delivery of inappropriate shocks is still possible. In this paper a singular case of T-wave oversensing is described, associated with a high number of inappropriate shocks and consequent rapid depletion of ICD battery. Persistent abnormalities of cellular repolarization following a first appropriate shock were believed to be at the origin of T-wave oversensing. J Med Cases. 2014;5(2):92-97 doi: http://dx.doi.org/10.14740/jmc1642w
No takes yet. Share an insight, caveat, or question.
Campana (2014) conducted a case report in T-wave oversensing and inappropriate ICD shocks (n=1). Implantable Cardioverter Defibrillator (ICD) was evaluated on Inappropriate shocks and battery depletion. An initial appropriate ICD shock induced persistent cellular repolarization abnormalities, resulting in T-wave oversensing that triggered 84 inappropriate shocks and rapid battery depletion.
Synapse has enriched 10 related papers. Consider them for comparative context: