Key result
A subcutaneous implantable cardioverter defibrillator resulted in inappropriate shocks due to T-wave oversensing and delayed therapy for slower ventricular tachycardia, necessitating explantation.
Why the study?
S-ICD has limitations including inappropriate shocks from oversensing and inability to treat VT below 170 bpm, which can both manifest in the same patient to prompt device explant.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
This case highlights the critical importance of S-ICD pre-implant vector screening and attention to VT cycle length to prevent inappropriate shocks and delayed therapies.
BACKGROUND A subcutaneous implantable cardioverter defibrillator (S-ICD) is preferred over a transvenous implantable cardioverter defibrillator (TV-ICD) in selected cases owing to a lower rate of lead-related complications such as infections and venous thrombosis. However, the S-ICD has its own limitations, including inappropriate shocks due to oversensed events, and the inability to treat ventricular tachycardia (VT) below a heart rate of 170 beats per minutes (bpm). We present a patient case which showed manifestations of both of these limitations, warranting explant of the device. CASE REPORT A 50-year-old man with a history of nonischemic cardiomyopathy and VT had a S-ICD placed at an outside facility. However, he continued to have VT despite on anti-arrhythmic drugs and required recurrent S-ICD shocks. Device interrogation showed that he was intermittently receiving appropriate shocks for slower VT (with a heart rate ranging from 150 bpm to 160 bpm) due to oversensing of T waves. However, treatment was delayed for other VT episodes owing to appropriate sensing and the patient's heart rate being below the lowest detection zone for S-ICD. Due to slower VT cycle length and frequent oversensed events, the S-ICD was ultimately replaced by a TV-ICD system. CONCLUSIONS This case report emphasizes the importance of S-ICD pre-implant vector screening and the need for paying attention to VT cycle length to prevent inappropriate device shocks and/or delayed therapies.
No takes yet. Share an insight, caveat, or question.
Dhawan et al. (2021) conducted a case report in Ventricular tachycardia (n=1). Subcutaneous implantable cardioverter defibrillator (S-ICD) vs. Transvenous implantable cardioverter defibrillator (TV-ICD) was evaluated on Clinical course. A subcutaneous implantable cardioverter defibrillator resulted in inappropriate shocks due to T-wave oversensing and delayed therapy for slower ventricular tachycardia, necessitating explantation.
Synapse has enriched 10 related papers. Consider them for comparative context: