Why the study?
The study was conducted to develop and validate mSASHA, a simultaneous cardiac T1 and T2 mapping technique with high accuracy and precision in a single breath-hold.
Does the mSASHA sequence improve the accuracy and precision of simultaneous myocardial T1 and T2 mapping compared to conventional mapping techniques?
Does the mSASHA sequence improve the accuracy and precision of simultaneous myocardial T1 and T2 mapping compared to conventional mapping techniques?
The mSASHA sequence provides accurate and precise simultaneous quantification of myocardial T1 and T2 in a single breath-hold, overcoming the limitations of conventional separate mapping sequences.
May enable efficient T1/T2 mapping; leaves open clinical adoption pending prospective validation.
Purpose: To develop and validate a multi-parametric SAturation-recovery single-SHot Acquisition (mSASHA) cardiac T1 and T2 mapping technique with high accuracy and precision in a single breath-hold. Methods: The mSASHA acquisition consists of 9 images in an 11 heartbeat breath-hold -- the first without preparation, 6 images with saturation recovery preparation, and 2 images with both saturation recovery and T2-preparation. T1 and T2 values were calculated using a 3-parameter model. mSASHA was validated in simulations and phantoms on a Siemens 3T Prisma scanner with comparison to a joint T1-T2 technique with a 4-parameter model. mSASHA values were compared to reference MOLLI, SASHA and T2p-bSSFP sequences in 10 healthy volunteers. Results: mSASHA had high accuracy compared to reference spin-echo measurements, with an average of -0.7+/-0.4% T1 error and -1.3+/-1.3% T2 error. mSASHA coefficient of variation (CoV) in phantoms for T1 was lower than MOLLI (0.7+/-0.1% vs 0.9+/-0.2%, p<0.01) and similar to reference T2p-bSSFP for T2 (1.4+/-0.6% vs 1.5+/-0.5%, p>0.05). In simulations, 3-parameter mSASHA fitting had higher precision than 4-parameter joint T1-T2 fitting for both T1 and T2. In-vivo myocardial mSASHA T1 was similar to conventional SASHA (1523+/-18 ms vs 1520+/-18 ms, p>0.05) with similar CoV to both MOLLI and SASHA (3.3+/-0.6% vs 3.1+/-0.6% and 3.3+/-0.5% respectively, p>0.05 for both). Myocardial mSASHA T2 values were 37.1+/-1.1 ms with similar precision to T2p-bSSFP (6.7+/-1.7% vs 6.0+/-1.6%, p>0.05). Conclusion: mSASHA provides high accuracy cardiac T1 and T2 quantification in a single breath-hold, with similar precision to reference MOLLI and linear T2p-bSSFP reference techniques.
No takes yet. Share an insight, caveat, or question.
Chow et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: