Key result
SMART1Map provided the best combination of true T1 accuracy, high repeatability, and low variability compared to SASHA and MOLLI in a phantom study.
Population
A phantom containing 22 samples of different T1s (200-1600 ms) and T1/T2 ratios (1.1 - 20.4)
Comparison
SMART1Map and SASHA T1 mapping techniques vs MOLLI T1 mapping technique
Design
Preclinical
Authors
Loading...
Phantom data favor SMART1Map for T1 mapping; hypothesis-generating for in vivo validation before clinical consideration.
In a phantom study, SMART1Map demonstrated superior accuracy, repeatability, and low variability for T1 mapping compared to MOLLI and SASHA.
Stainsby et al. (2014) studied Phantom study for myocardial T1 mapping (n=22). SMART1Map, SASHA, and MOLLI vs. Reference T1s from conventional single-point IR-spin echo was evaluated on Accuracy, repeatability, and pixel-wise variability of measured T1 values. SMART1Map provided the best combination of true T1 accuracy, high repeatability, and low variability compared to SASHA and MOLLI in a phantom study.