Does electroconvulsive therapy without ICD deactivation cause inappropriate discharges in patients with an ICD?
Does electroconvulsive therapy without ICD deactivation cause inappropriate discharges in patients with an ICD?
Answers are grounded in this paper and cite their sources.
In a patient with an ICD undergoing 70 ECT sessions without device deactivation, no inappropriate shocks occurred, suggesting the risk of inappropriate therapy may be lower than currently assumed.
Electroconvulsive therapy (ECT) is an effective treatment for severe depressive patients even in high-risk cardiac patients, such as those with an implantable cardioverter-defibrillator (ICD), but current practice guidelines recommend disabling all antitachycardia therapy before each ECT session. We describe a patient with severe psychotic depression who underwent a series of 70 ECT sessions without deactivation of the ICD. The figure shows continuous monitoring of the ICD with the programmer without significant ventricular oversensing. The electrical interference produced minimal noise levels with inappropriate detection of several ventricular events—asterisks [classified as ventricular sense events, ventricular fibrillation events, and other events classified as noise (VN)]. There were no detections of inappropriate VF episodes and, therefore, there were no inappropriate discharges during the course of 70 ECT sessions. The reported case suggests that even though ICD therapies should be temporarily deactivated, the risk of inappropriate therapy due to ECT seems to be low. Conflict if interest: none declared.
No takes yet. Share an insight, caveat, or question.
Anguera et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: